|
The investigation determined that a lower than expected vitros acetaminophen (acet) result was obtained from a single patient sample tested on a vitros 5600 integrated system.The result was lower than expected when compared to the initial result from the same patient sample.A definitive cause of the event could not be determined given the limited information provided.The customer did not provide baseline mean/sd values, package insert (pi) values, and/or randox peer values to the ortho tsc when requested.This information would be used to evaluate the customers historic vitros acet quality control (qc) results leading up to the day of the event.As this information was not provided, historic qc results were unable to be evaluated.Therefore, a vitros acet slide lot 2918-0077-1305 performance issue cannot be completely ruled out as contributing to the event.Continual tracking and trending of complaints has not identified any signals that would point to a potential systemic performance issue with vitros acet slide lot 2918-0077-1305.As of the date of writing this report, no vitros alkp diagnostic precision testing has been performed; therefore, an instrument related issue cannot be ruled out as contributing to the event.Pre-analytical sample mix-up was not a likely contributor to the event, as a review of the sample viscosities and zee-height measurements showed consistent results between the initial and repeat of sample 1, indicating that the same sample was processed.It is unknown if the customer was following the collection device manufacturers specifications for centrifugation, therefore, improper pre-analytical sample processing could not be ruled out as contributing to the event.It is possible that cellular debris, due to poor sample preparation, was present in the affected sample, although this could not be confirmed.
|