| Catalog Number |
51-103140 |
| Medical Device Problem Codes |
Delivered as Unsterile Product (1421); Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the stem punctured the sterile packaging.There was no known impact or consequences to the patient.It was reported no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: d9; g3; h2; h3; h6.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister and pouch).Sterility has been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.The root cause of the reported event can be attributed to transit damage and a packaging design issue.Complaint is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|