U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG PLEXA PROMRI S 60; ICD LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG PLEXA PROMRI S 60; ICD LEAD Back to Search Results
Model Number 402265
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Cardiogenic Shock (2262)
Date of Event 11/24/2025
Type of Reportable Event Serious Injury
Event or Problem Description
System was explanted due to infection.Should additional information become available, this file will be updated.
 
Additional Manufacturer Narrative
Combination product: yes.An infection was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally, an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the infection was not device related.
 
Additional Manufacturer Narrative
Combination product: yes.16-dec-2025 additional information received.Patient was admitted to facility (b)(6) 2025, a picc procedure occurred (b)(6) 2025 with post-procedure diagnosis of cardiogenic shock, then system was explanted on (b)(6) 2025.An infection was observed following the implantation of this biotronik device.The sterilization process was investigated.The validated process assures that all sterilization parameters, such as gas concentration, temperature, humidity, etc., are within its specified ranges for each distributed device.Additionally, an analysis of validated microbiological indicators is performed after every sterilization procedure as evidence of successful completion of the sterilization process.Review of the biotronik complaint database did not reveal any changes regarding the trend for this type of incident.In summary, the infection was not device related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLEXA PROMRI S 60
Common Device Name
ICD LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key23821963
Report Number1028232-2025-06165
Device Sequence Number9974536
Product Code NVY
UDI-Device Identifier04035479140014
UDI-Public(01)04035479140014(17)220731
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date07/31/2022
Device Model Number402265
Device Catalogue Number402265
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2025
Supplement Date Received by Manufacturer12/16/2025
Initial Report FDA Received Date12/16/2025
Supplement Report FDA Received Date12/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/27/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age52 YR
Patient SexMale
-
-