D10 concomitant product: egiauxl endogia ultra univ xl stapler (lot#: p5h0932).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
It was reported that during a hemicolectomy, the stapler was opened and a purple reload was loaded, cycled, and closed.When the surgeon attempted to insert it into the port, the jaws would not open.The device was removed from the patient and remained unable to be opened.After the case, the scrub nurse managed to open the stapler but noted a ¿grindy¿ sensation when taking the reload off.The device was replaced with a new stapler and reload.There was no patient injury.
|
|
Additional information: d1, d4 (model #, catalog #,expiration date, lot #, unique identifier (udi) #), d8, d9, g3, g4 (pma/510(k) #), h3, h4, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the jaws would not open.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|