• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OSTEOREMEDIES LLC REMEDY SPECTRUM GV HIP SPACER; HIP JOINT FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OSTEOREMEDIES LLC REMEDY SPECTRUM GV HIP SPACER; HIP JOINT FEMORAL Back to Search Results
Medical Device Problem Codes Loss of or Failure to Bond (1068); Fracture (1260)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The article discussed that the patients were weight bearing as tolerated.A review of the product labeling identified the following precaution" the remedy spectrum® gv hip spacer may be compromised in an overweight or obese patient and/or one who does not limit the amount of activity and weight placed on the hip.Always use the largest component size possible to ensure ideal performance.It is essential that the patient use mobility-assisted devices (e.G., crutches, walker, cane) during the implantation period." the article did not discuss the implantation technique for implantation of the stem into the canal.The product labeling identifies the following warning "the remedy spectrum® gv hip spacer must be proximally cemented with gentamicin and vancomycin loaded cement.The use of bone cement is compulsory to avoid rotation and to limit the risk of dislocation or spacer loosening." the article demonstrated that additional cement can be beneficial in addition to the use of the monomer.The product labeling contains the following note: "the remaining offset space and threads of the stem, up to the femoral head, must be filled with gentamicin and vancomycin loaded bone cement.".
 
Event or Problem Description
Journal of orthopaedic experience and innovation vol.6, issue2, 2025 included reserach article "stability of osteoremedies remedy hip spacer for two-stage revisions of prosthetic joing infections: single center analysis" included two patients who required spacer replacment.Information from the article: in this series, the osteoremedies® second-generation pre-molded antibiotic hip spacer proved to be a convenient and effective option during two-stage revisions for managing pjis, though it demonstrated a small risk of implant fracture and dislocation.This spacer demonstrated only 1 fracture and 1 dislocation of the 16 patients who used both stem and head pieces.Of the 16 patients who used both femoral components, a spacer fracture occurred in one case (6.25%).This patient underwent a spacer exchange.Based on initial post-operative x-rays, the degree of threading of the femoral neck into the head was 6 threads, but at the time of fracture, it had reduced to only 2 threads.This patient did not receive additional bone cement, and their bmi was 34.75 kg/m².When comparing this patient who experienced a spacer fracture to another in our cohort with the same thread count of 6 who did not fracture, the key differences between the two patients were the use of supplemental bone cement and bmi of 22.7 kg/m².Of the 16 patients who used both femoral components, a spacer dislocation occurred in 1 case (6.25%).In this case, the spacer was exchanged with a different spacer.The degree of threading of the femoral neck into the head was 4 threads.For this patient, additional bone cement was not used, their bmi was 38.02 kg/m², and the patient reported a fall resulting in twisting of the affected limb.When comparing the demographics of the patient who experienced spacer dislocation to the other patient in our cohort with 4 threads who did not dislocate, the only differences were the use of supplemental cement, bmi 24.4 kg/m², and presence of trauma.Among the 16 patients who utilized both components, the femoral head was screwed onto the stem with an average of 7.75 (range: 4-9) threads.The average change in distance between the femoral head and stem components from post-spacer insertion to the most recent followup was 0.15 mm (range: -0.5 to 2.93 mm).Of the 10 patients who used supplemental bone cement around the junction between the femoral stem and head, the average change in distance was -0.15 mm.Of the 6 patients who did not use supplemental cement, the average change in distance was 0.78 mm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMEDY SPECTRUM GV HIP SPACER
Common Device Name
HIP JOINT FEMORAL
Manufacturer (Section D)
OSTEOREMEDIES LLC
6800 poplar ave
suite 120
memphis TN 38119
Manufacturer Contact
chris hughes
6800 poplar ave
memphis, TN 38119
9017340445
MDR Report Key23824711
Report Number3010537287-2025-00002
Device Sequence Number5093233
Product Code KWL
UDI-Device Identifier00855195006524
UDI-Public00855195006524
Combination Product (Y/N)Y
PMA/510(K) Number
K192885
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Literature
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2025
Initial Report FDA Received Date12/16/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-