| Catalog Number |
212480 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Application Program Problem: Parameter Calculation Error (1449)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following information was provided: the sagittal attachment from the mako saw vibrates abnormally.Update from mps on (b)(6) 2025: all i can say is that we were able to make the incisions, but the surgeon felt that there was an up-and-down oscillation in addition to the normal oscillation.In addition, it made a different noise than usual.Unfortunately, i can't say any more than that.I hope this has been helpful to you nonetheless.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding mechanical failure involving a mako saw attachment was reported.The event was not confirmed because the product was not available for inspection.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Review of the device history records indicate devices were manufactured and accepted into final stock on with no relevant reported discrepancies.Complaint history review: there have been other complaints with similar event(s) for the lot referenced.Conclusions: the alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|