| Model Number |
SUTURE ANCHOR, PUSHLOCK 3.5 X 19.5MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2025, it was reported by an arthrex's employee via an email that for 2 units of ar-1926ps, suture anchor, pushlock® 3.5 x 19.5 mm, the first unit of ar-1926ps broke and the second unit of ar-1926ps could not be fixed during insertion.This was detected during an ankle ligament reconstruction surgery on (b)(6) 2025.The procedure was completed successfully by using a 3rd ar-1926ps.There were no patient harm and no secondary procedure needed.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
|
| |
|