| Model Number |
1400300-08 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the information received, the investigation determined that the reported complaints appear to be related to operational context.There was no leak noted to the balloon during preparation or during the first inflation, which suggests a product quality issue did not contribute to the reported difficulties.It was reported by the account that the balloon dilatation catheter (bdc) interacted with the moderately tortuous and moderately calcified anatomy resulting in the reported difficulty advancing the device and subsequently the balloon became damaged and ruptured during the second inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported the 3.00 x 8mm nc trek neo rx was to be used in the left anterior descending (lad) lesion with resistance between the balloon and the lesion.However, the nc trek ruptured at 8 atmospheres (atm) during the second inflation.Therefore, the nc trek was removed and another nc trek balloon was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|