| Model Number |
SCORPION NEEDLE, KNEE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament surgery the tip of the needle broke and was stuck inside the scorpion.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The reported event was confirmed.One unpacked ar-12990 with batch 15186526 was received with a broken fragment of the needle stuck inside.Remaining parts of the needle have not been received, and no further functional testing was possible due to the observed condition.This type of event is most likely resulting from improper device handling.Dfu-0392-sub-eo warns against the usages listed to help avoid needle breakage and potential patient injury.
|
| |
|