Catalog Number 517750
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
According to the available information the device was explanted and replaced due to lack of fluid in system.
 
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, non-conforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no non-conforming reports or capas that were confirmed to be associated.The device was not returned for evaluation.Without the benefit of analyzing the device, quality cannot comment on the condition of the device.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
 
Event or Problem Description
According to the available information the device was explanted and replaced due to lack of fluid in system.
 
Additional Manufacturer Narrative
Titan touch pump, cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.A separation was noted on the longer exhaust tubing of the pump.This was a site of leakage.The surfaces appear to be rough and irregular, indicating stress was exerted.A failure of the back pressure and inlet valve tests were noted with the pump as the balls in the pump failed to seat properly.It was determined that foreign material was noted in the spring/ball and seat component of the pump.Abrasion was noted on the exhaust tubing of cylinder 1.No functional abnormalities were noted with cylinder 1 or cylinder 2.No functional abnormalities were noted with the reservoir.Based on examination of the returned product, it was concluded that the rough and irregular surfaces associated with the separation indicate that sufficient stress was exerted on the longer exhaust tube of the pump to separate the site while in-vivo.A separation of this type could then allow the loss of fluid, making the device inoperable.Review of the returned pump was conducted.During this review, it was concluded that the foreign material noted in the spring/ball and seat component of the pump resulted in the failure of the back pressure and inlet valve testing.Because these components were released according to manufacturing and quality control procedures, it was concluded that the foreign material observed most likely entered the system after the device packaging was opened.Failure of this test was not associated with the cause of the reported device malfunction.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Common Device Name
PENILE PROSTHESIS, INFLATABLE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
MINNEAPOLIS
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key23828141
Report Number2125050-2025-01936
Device Sequence Number5068148
Product Code FHW
UDI-Device Identifier05708932533966
UDI-Public5708932533966
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/03/2024
Device Catalogue Number517750
Device Lot Number6837855_5177502400
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/17/2025
Supplement Report FDA Received Date06/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexUnknown
-
-