Catalog Number 205548000
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional manufacturer narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during sterile processing staff identified deep scratches and grooves on the modular endo head trials.The event did not occur during surgery, and there was no patient involvement.No additional information could be provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: it was reported that the modular endo head trials identified by staff as needing replaced due to deep scratches and grooves.Did not happen in surgery the product was returned to depuy synthes for evaluation.Visual analysis of the returned device revealed that the s/c & mod cath trl had scratches on the surface.This type of damage is consistent with other tools and hard edges coming in contact with the device.Properly handling and attention to the approved use of the device diminishes the risk of failure.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the s/c & mod cath trl would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (af0408) provided is not a valid finished goods lot number.H11 additional narrative: corrected: h3.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S/C & MOD CATH TRL 48/28
Common Device Name
HIP INSTRUMENTS : ACETABULAR TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23828543
Report Number1818910-2025-22064
Device Sequence Number9288102
Product Code HWT
UDI-Device Identifier10603295084655
UDI-Public(01)10603295084655(10)AF0408
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number205548000
Device Lot NumberAF0408
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/05/2025
Supplement Date Received by Manufacturer01/05/2026
01/09/2026
Initial Report FDA Received Date12/17/2025
Supplement Report FDA Received Date01/06/2026
01/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/15/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1