• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS) BD MAX¿ ENTERIC VIRAL PANEL; GASTROINTESTINAL PATHOGEN PANEL MULTIPLEX NUCLEIC ACID-BASED ASSAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS) BD MAX¿ ENTERIC VIRAL PANEL; GASTROINTESTINAL PATHOGEN PANEL MULTIPLEX NUCLEIC ACID-BASED ASSAY SYSTEM Back to Search Results
Catalog Number 443985
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D.2.Additional medical device type: pch.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that during use of bd max¿ enteric viral panel, a false positive norovirus patient result with a week pcr curve was obtained.There was no report of patient impact.
 
Additional Manufacturer Narrative
Investigation summary: the complaint investigation for discrepant results with the bd max¿ enteric viral panel kit (ref.(b)(4)) lot 5021181 was performed by the review of manufacturing records, customer¿s data analysis and verification of complaints history.The review of quality control records of bd max¿ enteric viral panel kit lot 5021181 revealed no anomalies that could explain the customer¿s discrepant results.Customer reported false positive results for norovirus (nov) with the bd max¿ enteric viral panel.According to customer, there was no real amplification or no sigmoidal curves, and sometimes bubbles could be observed in cartridges.Customer provided run files (pdf) of run runs 488 from bd max¿ instrument ct3307 for investigation.Samples in run 488 in positions a4 and a5 were identified as problematic by the customer.Manual pcr curve adjudication revealed step dislocations in multiple channels were observed, and it is unlikely these positive results are true amplifications.These results could be related to the bubbles the customer mentioned having observed in the cartridges, since multiple channels are impacted at the same cycle.Overall, bd was unable to find the exact cause of the customer¿s positive results, but based on the investigation, no reagents issue is suspected.There is no indication of a reagent issue based on the analysis of the complaints received for discrepant results on bd max¿ enteric viral panel assay lot 5021181.The root cause was not identified.Bd cannot confirm the complaint based on the investigation that was performed.Bd did not initiate a corrective and preventive action (capa) since no new hazard was identified.
 
Event or Problem Description
Report 2 of 2: it was reported that during use of bd max¿ enteric viral panel, a false positive norovirus patient result with a week pcr curve was obtained.There was no report of patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAX¿ ENTERIC VIRAL PANEL
Common Device Name
GASTROINTESTINAL PATHOGEN PANEL MULTIPLEX NUCLEIC ACID-BASED ASSAY SYSTEM
Manufacturer (Section D)
GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS)
2555 blv. du parc techn
quebec
Manufacturer (Section G)
GENEOHM SCIENCES CANADA, INC. (BD DIAGNOSTICS)
2555 blv. du parc techn
quebec
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key23828881
Report Number3007420875-2025-00266
Device Sequence Number9267882
Product Code OOI
UDI-Device Identifier00382904439859
UDI-Public(01)00382904439859(17)260808(10)5021181
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K220607
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/08/2026
Device Catalogue Number443985
Device Lot Number5021181
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2025
Supplement Date Received by Manufacturer02/02/2026
Initial Report FDA Received Date12/17/2025
Supplement Report FDA Received Date02/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-