• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 2.0MM DRILL BIT WITH DEPTH MARK/QC/140MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.0MM DRILL BIT WITH DEPTH MARK/QC/140MM Back to Search Results
Catalog Number 323.062
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that drill bit ø2 w/double marking l140/115 3, drill bit ø2/1.15 cann l150/48 3flute f/, and k-wire ø1 l150 sst were involved in an intraoperative event.During the procedure, shortages of a 2.0 drill bit and a 1.0 k-needle were identified upon opening the equipment.Later, the 2.0 drill bit broke, leaving its tip embedded in the patient's bone.The fragment could not be removed and was intentionally left in place, as it was determined there would be no adverse consequence.The remaining portion of the drill bit was removed, and a replacement drill bit was used to complete the surgery without delay or discomfort to the specialist.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d10 (concomitant).H3, h6: photo investigation: the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that '323.062, drill bit ø2 w/double marking l140/115 3¿ had broken.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.The allegation of embedded device can not be confirmed as no patient x-rays were provided to evaluate the condition.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the drill bit ø2 w/double marking l140/115 3 would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to component failure and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): a dhr evaluation was performed for the finished device.Product code: 323.062.Lot number: 76971p5.It was electronically reviewed and no nonconformances/manufacturing irregularities were identified during the manufacturing process.The product was released on: 07-august-2025.Manufacturing site: jabil bettlach.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM DRILL BIT WITH DEPTH MARK/QC/140MM
Common Device Name
DRILL BIT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key23829243
Report Number8030965-2025-12695
Device Sequence Number13083372
Product Code KTW
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number323.062
Device Lot Number76971P5
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/14/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/01/2025
Supplement Date Received by Manufacturer12/30/2025
09/14/2026
09/23/2026
09/24/2026
Initial Report FDA Received Date12/17/2025
Supplement Report FDA Received Date12/31/2025
09/16/2026
09/24/2026
09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/07/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DRILL BIT Ø2/1.15 CANN L150/48 3FLUTE F/; K-WIRE Ø1 L150 SST
Patient SexMale
-
-