• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SUPER SHEATH INTRODUCER SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SUPER SHEATH INTRODUCER SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number 28211
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the device broke off, requiring additional intervention for removal.A 6 fr x 11 cm.035 super sheath introducer sheath was selected for placement.During the procedure, there was no resistance noted when the super sheath device was advanced in the vessel.However, upon completion of the procedure, it was noted that the device broke under fluoroscopic visualization.An attempt was made to extract the foreign body using a non-boston scientific (non-bsc) snare, without success.The detached portion of the device remained in the common femoral artery.A non-bsc 7 fr femoral introducer was then replaced, followed by an 8 fr femoral introducer.The system was mounted on a 7 fr jr 3.5 guide catheter, also without success.Additionally, a non-bsc loop-shaped angioplasty guide was used, but after multiple failed attempts to retrieve the device, the procedure was terminated due to the high risk of associated vascular complications.The patient underwent open surgery to explore and retrieved the introducer sheath.The cardiac catheterization procedure was completed.The patient fully recovered post-procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPER SHEATH INTRODUCER SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
MEDIKIT COMPANY LIMITED
13-2 yushima 1-chome
bunkyo-ku tokyo 113-0 034
JA   113-0034
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key23829396
Report Number2124215-2025-87294
Device Sequence Number13083494
Product Code DYB
UDI-Device Identifier14543527202449
UDI-Public(01)14543527202449(17)260131(10)23B14B7
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/30/2026
Device Model Number28211
Device Catalogue Number28211
Device Lot Number23B14B7
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2025
Initial Report FDA Received Date12/17/2025
Date Device Manufactured02/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE - ASAHI SION BLUE; SNARE - MULTISNARE SYSTEM IN 4 AND 6 FR
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexFemale
-
-