• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RXSIGHT, INC. LIGHT ADJUSTABLE LENS+ (LAL+)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RXSIGHT, INC. LIGHT ADJUSTABLE LENS+ (LAL+) Back to Search Results
Model Number 60007
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 11/18/2025
Type of Reportable Event Serious Injury
Event or Problem Description
A site reported to rxsight that a patient's light adjustable lens+ (lal+, sn (b)(6), +18.5d) was explanted due to the patient's dissatisfaction with the chosen refractive target and a new lal+ (sn (b)(6), +18.5d) implanted as a replacement.
 
Additional Manufacturer Narrative
The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.Manufacturer reference #: (b)(4).
 
Additional Manufacturer Narrative
This supplemental report is submitted to provide the results of the investigation and device evaluation.Updates to sections h3 and h6.The explanted lal+ was returned to rxsight.A visual inspection found the lens cut into 3 pieces and the haptics missing, all attributed to the lens explantation procedure.Other than this damage, no anomalies were noted on the lal+.Analysis of information provided by the surgeon noted that this lens (os) had been targeted to near vision while the opposing lens (od - not explanted) had been targeted for plano; however, the patient was unable to neuroadapt to the resulting monovision.Based on the available information, the patient's inability to adapt to monovision is the likely cause of the reported patient dissatisfaction with chosen refractive target and lens explant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHT ADJUSTABLE LENS+ (LAL+)
Common Device Name
LIGHT ADJUSTABLE LENS+ (LAL+)
Manufacturer (Section D)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer (Section G)
RXSIGHT, INC.
100 columbia
aliso viejo CA 92656
Manufacturer Contact
jim zhang
100 columbia
aliso viejo, CA 92656
9492736163-159
MDR Report Key23829709
Report Number3012712027-2025-00433
Device Sequence Number4992843
Product Code PZK
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P160055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number60007
Device Catalogue Number60007
Device Lot NumberL08-006223
Date Returned to Manufacturer12/03/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2025
Supplement Date Received by Manufacturer05/26/2026
Initial Report FDA Received Date12/17/2025
Supplement Report FDA Received Date06/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexFemale
-
-