| Model Number |
60007 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
11/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A site reported to rxsight that a patient's light adjustable lens+ (lal+, sn (b)(6), +18.5d) was explanted due to the patient's dissatisfaction with the chosen refractive target and a new lal+ (sn (b)(6), +18.5d) implanted as a replacement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record for the manufacturing lot was reviewed and there were no discrepancies or unusual findings.Manufacturer reference #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is submitted to provide the results of the investigation and device evaluation.Updates to sections h3 and h6.The explanted lal+ was returned to rxsight.A visual inspection found the lens cut into 3 pieces and the haptics missing, all attributed to the lens explantation procedure.Other than this damage, no anomalies were noted on the lal+.Analysis of information provided by the surgeon noted that this lens (os) had been targeted to near vision while the opposing lens (od - not explanted) had been targeted for plano; however, the patient was unable to neuroadapt to the resulting monovision.Based on the available information, the patient's inability to adapt to monovision is the likely cause of the reported patient dissatisfaction with chosen refractive target and lens explant.
|
| |
|