| Catalog Number |
151820606 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total knee arthroplasty (tka) with a cruciate-retaining femoral multi-size polyethylene, the surgeon impacted the polyethylene insert and felt the locking mechanism did not engage correctly.The surgeon removed and reinserted the insert several times and opened a second and then a third polyethylene with the same issue.After a fourth polyethylene was opened and still did not satisfy the surgeon, a stemmed tibial component was opened in case tray removal was required and to evaluate whether the tibial tray might be causing the locking failure; the team attempted to engage the polyethylene and tray on the back table.The surgeon was not convinced that changing trays would resolve the problem and ultimately implanted a 6x7 left multi-size polyethylene, although the surgeon remained not fully satisfied at the conclusion of the case.No further information is available.Doe: (b)(6) 2025.Affected side: left knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition; therefore, it has been determined that no corrective and preventive action is required at this time.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
|
| |
|
Search Alerts/Recalls
|