STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
| Catalog Number |
5530-G-309 |
| Medical Device Problem Codes |
Degraded (1153); Fracture (1260); Unstable (1667)
|
| Health Effect - Clinical Codes |
Pain (1994); Joint Laxity (4526)
|
| Date of Event |
12/08/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The following was reported: "revision of size 3 triathlon poly.Patient presented with pain, some clinical instability reported.Can¿t see item number or lot number to confirm poly size 3 no original operation info found.".
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event:
an event regarding wear and crack/fracture involving triathlon insert was reported.The event was confirmed via evaluation of the returned device.Method & results:
-device evaluation and results: visual inspection of the returned device indicated that the insert is worn and part of the articulating surface has cracked.The damage present is consistent with delamination and material loss.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.A small portion of the posterior edge has fractured.The yellow discoloration observed on the tibial insert is consistent with the absorption of synovial fluid by the device.-clinician review: no medical records were received for review with a clinical consultant.-device history review: review of the product history records indicate the reported device was manufactured with no reported relevant discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion:
visual inspection of the returned device indicated that the insert is worn and part of the articulating surface has cracked.The damage present is consistent with delamination and material loss.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.A small portion of the posterior edge has fractured.The yellow discoloration observed on the tibial insert is consistent with the absorption of synovial fluid by the device.No further investigation is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|