Catalog Number 7757-2MP65/B
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow-up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that during a power injection and the hub from the catheter broke off.No reported injury.
 
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.Nonconformances of this nature are monitored by the operation team.This complaint will be used to trend for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORMA®.
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
calle, sor juana ines de la cr
roberto de la madrid
tijuana, b.c. 22630
MX  22630
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC.
calle, sor juana ines de la cr
roberto de la madrid
tijuana, b.c. 22630
MX   22630
Manufacturer Contact
adrian cole
1600 merit parkway
south jordan, UT 84095
MDR Report Key23831251
Report Number1721504-2025-00746
Device Sequence Number18399719
Product Code DQO
UDI-Device Identifier00884450766507
UDI-Public00884450766507
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K943739
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number7757-2MP65/B
Device Lot NumberI3201007S1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2025
Supplement Date Received by Manufacturer01/02/2026
Initial Report FDA Received Date12/17/2025
Supplement Report FDA Received Date01/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-