| Model Number |
VTICM5_12.6(-12.5/3.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/03/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - 4581: rotation.H6 - work order search: no additional similar complaint type events within associated lots were found.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation.Due to lens rotation, the lens was repositioned.This did not resolve the problem.On another day the lens was exchanged for a same model and length lens.The problem was resolved.Cause of the event is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - the reporter indicated that following an implantable collamer lens (icl) implantation procedure, the patient experienced lens rotation.Due to lens rotation, the lens was repositioned.This did not resolve the problem.On another day the lens was exchanged for a same model, length but change in axis lens.The problem was resolved.Cause of the event is unknown.H3 - device evaluation: the lens was returned dry in a microcentrifuge tube.Visual inspection found the lens optic torn and the haptic broken.Dimensional inspection found the lens to be within specifications.Claim # (b)(4).
|
| |
|