MAUDE Adverse Event Report: SYNTHES GMBH DRILL BIT Ø2.5 CALIBR L300 W/QUICK-COUPL; BIT, DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH DRILL BIT Ø2.5 CALIBR L300 W/QUICK-COUPL; BIT, DRILL Back to Search Results
Catalog Number 324.210
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2025, the patient underwent surgery for pelvic fractures with lpp.During surgery, the surgeon used the two drill bits, but neither were sharp enough nor could not be used.Another drill bit was used for the rest of the procedure.The surgery was completed successfully with no surgical delay.No further information is available.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d9, d10 (concomitant).The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1 (manufacturing site name).H3, h4, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device found blunting/rounding of cutting edges.The device shows signs of repetitive use.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the drill bit ø2.5 calibr l300 w/quick-coupl would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): part: 324.210.Lot: 1l90819.Manufacturing site: werk bettlach.Supplier: na.Release to warehouse date: 30 oct 2018.Expiration date; na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL BIT Ø2.5 CALIBR L300 W/QUICK-COUPL
Common Device Name
BIT, DRILL
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach CO 2544
SZ   2544
Manufacturer Contact
elba bello
1302 wrights lane east
west chester, PA 19380
9083863534
MDR Report Key23831464
Report Number8030965-2025-12711
Device Sequence Number9268441
Product Code HTW
UDI-Device Identifier07611819796973
UDI-Public(01)07611819796973(10)1L90819
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number324.210
Device Lot Number1L90819
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/01/2025
Supplement Date Received by Manufacturer01/14/2026
01/30/2026
Initial Report FDA Received Date12/17/2025
Supplement Report FDA Received Date01/15/2026
02/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DRILL BIT Ø2.5 QC L170 CALIBRATION 80.
Patient SexUnknown
-
-