| Model Number |
VIEW |
| Medical Device Problem Code |
Therapeutic or Diagnostic Output Failure (3023)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported there was an error resulting in a loss of mechanical ventilation during service.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Ge healthcare investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Block a: no report of patient involvement.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.
|
| |
|
Additional Manufacturer Narrative
|
|
The distributor performed a checkout of the equipment and confirmed the reported complaint.The enhanced sensor interface board (esib) was replaced to correct the reported issue.
|
| |
|