Model Number 97021HS
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation conclusion: the customer's complaint was not replicated during in-house testing with retained devices of lot t16083rn.No issues with recovery were observed and the product performed as expected.Manufacturing batch records for the lot were reviewed and found that the lot met release specifications.Unable to determine the root cause of the complaint.Based on the information available, there is no indication of a product deficiency and no corrective action is required.
 
Event or Problem Description
Customer reported poor correlation between triage and architect for 8 patients.No patient symptoms, treatments or diagnoses were available.Treatment was not based on triage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIAGE CARDIAC TNI PANEL
Common Device Name
TRIAGE CARDIAC TNI PANEL
Manufacturer (Section D)
QUIDEL CARDIOVASCULAR INC.
9975 summers ridge road
san diego CA 92121
Manufacturer Contact
chelsea lindgren
9975 summers ridge road
san diego, CA 92121
MDR Report Key23834711
Report Number3013982035-2025-00037
Device Sequence Number5067872
Product Code MMI
UDI-Device Identifier30014613337935
UDI-Public(01)30014613337935(10)T16083RN(17)260517(11)250903
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K030286
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2026
Device Model Number97021HS
Device Lot NumberT16083RN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2025
Initial Report FDA Received Date12/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/03/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-