EDWARDS LIFESCIENCES DR SWAN-GANZ PACING TD CATHETER; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number D205F7
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Tricuspid Valve Insufficiency/ Regurgitation (4453)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available for return as it was discarded by the facility.However, a supplemental report will be forthcoming when the enginering evaluation is completed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event or Problem Description
It was reported in an academic article that during use of the swan-ganz catheter the patient's tricuspid valve chordae tendineae ruptured.A 55 year old male underwent total arch replacement for stanford type b aortic dissection.Preoperatively, cardiac function was preserved and valvular regurgitation was trivial.Immediately after surgery, chest radiography revealed that the pac was kinked within the pulmonary artery approximately 6 cm from the tip.The catheter was withdrawn under radiographic monitoring, but the kink was not resolved.It was further withdrawn to a depth of 15 cm while observing the pressure waveform, then advanced again into the pulmonary artery after balloon inflation.Subsequent radiography confirmed that the kink had been corrected.The postoperative course was initially uneventful however, on postoperative day 9, body weight gain was noted, and echocardiography revealed new-onset severe tricuspid regurgitation.On postoperative day 15, tricuspid valve repair was performed.Initially, per the article, it is highly likely that when the kinked pa catheter was withdrawn into the right atrium, the kinked portion became caught on the tricuspid valve chordae tendineae, resulting in chordal injiury.Per information received from a bd apm sales rep, the event of tricuspid valve chordae tendineae was a procedural complicataion and not due to device malfunction.The patient was discharged after the tricuspid valvuloplasty without any major issues.
 
Additional Manufacturer Narrative
The device was not returned for product evaluation as it was discarded.Without return of the unit, it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.The device history record review cannot be completed as the lot number was not provided.An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.As no device was returned, a product non-conformance or device failure could not be confirmed and no root cause could be determined.In this case, there was also no allegation or indication a device malfunction contributed to this adverse event.Investigation results indicate root cause is not likely to be related to the swan-ganz device.The ifu has been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.As part of the manufacturing process 100% of the units go through a tip inspection and leak and flow test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ PACING TD CATHETER
Common Device Name
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer (Section G)
BD APM
17200 laguna canyon rd
irvine CA 92618
Manufacturer Contact
erika bonilla
1 edwards way
irvine, CA 92614
MDR Report Key23835474
Report Number2015691-2025-10298
Device Sequence Number18399757
Product Code LDF
UDI-Device Identifier00690103197334
UDI-Public00690103197334
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K791183
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Literature,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberD205F7
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/05/2025
Supplement Date Received by Manufacturer02/04/2026
Initial Report FDA Received Date12/17/2025
Supplement Report FDA Received Date02/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age55 YR
Patient SexMale
-
-