• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. HEADWAY; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. HEADWAY; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number MC162167S-AL
Medical Device Problem Codes Entrapment of Device (1212); Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/26/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on a review of the device¿s risk documentation, the reported event did not indicate there were the presence of any potential or new manufacturing, design, quality, or other systemic issues, or non-conformance's.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform an analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformance's associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): warnings the microcatheter should be advanced or manipulated under fluoroscopic guidance.Do not advance or withdraw the device when excessive resistance is met until the cause of resistance is determined.Infusion pressure should not exceed 700 psi to avoid potential rupture of the microcatheter.Precautions exercise care in handling the microcatheter to reduce the chance of accidental damage.With the exception of dimethyl sulfoxide (dmso), use of organic solvents may damage the microcatheter and/or coating on the surface.Potential complications include, but are not limited to: vessel or aneurysm perforation, vasospasm, hematoma at the site of entry, embolism, ischemia, intracerebral/intracranial hemorrhage, pseudoaneurysm, seizure, stroke, infection, vessel dissection, thrombus formation, and death.To reduce the risk of damage or separation of the device, avoid repeated bending at the same point of the microcatheter.Take precaution when manipulating the microcatheter in tortuous vasculature to avoid damage to the microcatheter.Avoid advancing or withdrawal against resistance until the cause of resistance is determined.Directions for use carefully insert the distal section of the guidewire into the microcatheter hub.Advance the guidewire until the distal tip is near the distal end of the microcatheter.Slip the torque device over the proximal end of the guidewire to the desired location.A guiding catheter is placed into the appropriate vessel and the microcatheter/guidewire assembly is then advanced through the guiding catheter to the target vessel or vascular lesion.Set up a continuous flush of heparinized saline by connecting rhvs with pressurized flush solution lines to the hub of the guiding catheter and microcatheter.Carefully advance the microcatheter/guidewire to the guiding catheter distal tip.During navigation in the vasculature, advance the guidewire a short distance, then advance the microcatheter over the guidewire and repeat until the desired site is reached.The proximal portion of the microcatheter does not have the hydrophilic surface and may encounter resistance when this section is advanced through the rhv.Once the desired location has been reached, the guidewire is removed from the microcatheter.The diagnostic or therapeutic agent(s) are then prepared for delivery through the microcatheter.Warning: do not exceed the maximum recommended infusion pressure of 700 psi.Between uses, rinse the microcatheter in a basin of heparinized saline and wipe it gently with sterile, wet gauze and place in a basin of heparinized saline or a flushed dispenser tube to keep the hydrophilic surface wet until use.The most distal portion of the catheter remains the patient and the remainder of the device was discarded by the user facility.The physical device was not available for evaluation to investigate if a condition existed that would have contributed to the reported event.Supplemental imaging was also not provided for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation further examine the cause of the reported event.
 
Event or Problem Description
It was reported that during a left side mma embolization with liquid embolic, a headway duo microcatheter became entrapped in liquid embolic.The physician left a portion of the catheter in the embolic cast and removed the other portion.The patient is reported to be doing well and is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEADWAY
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key23835946
Report Number2032493-2025-90720
Device Sequence Number20103849
Product Code DQO
UDI-Device Identifier00842429117415
UDI-Public00842429117415
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K120917
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC162167S-AL
Device Catalogue NumberMC162167S
Device Lot Number0001198188
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2025
Initial Report FDA Received Date12/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TRUFILL LIQUID EMBOLIC.
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-