• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION SD ALARIS SYSTEM; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SD ALARIS SYSTEM; PUMP, INFUSION Back to Search Results
Model Number 8015
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record, complaint history review, and risk review on failure modes are performed on each device with reportable malfunction(s) along with other methods of investigation as coded in section h6 of this mdr report.This report lists imdrf annex a, c, d, and g codes that are reportable malfunctions observed on the device during servicing.Per 803.52(f)(11)(iii) the information provided was obtained from servicing activities performed on the device.There were no additional details obtainable or provided at the time of service.
 
Event or Problem Description
It was reported that the device had not giving numerical data after the battery has been reconditioned, stuck in discharge status, replaced with new battery and same condition.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS SYSTEM
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
brett wilko
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key23836692
Report Number2016493-2025-144499
Device Sequence Number14725723
Product Code FRN
UDI-Device Identifier10885403519215
UDI-Public(01)10885403519215
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K211218
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 11/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number8015
Device Catalogue Number8015 ALARIS PCU 1.5 MODULE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/24/2025
Initial Report FDA Received Date12/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/16/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-