| Catalog Number |
368650 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Difficult or Delayed Activation (2577)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3.Date of event is unknown.The date received by manufacturer has been used for this field.D4.Medical device expiration date: unknown.H4.Device manufacture date: unknown.G5: additional pma / 510(k)#: k982541.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that when using one (1) bd vacutainer® eclipse¿ blood collection needle pre-attached holder, the safety mechanism separated, and tube push off from the non-patient end of the cannula occurred.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd did not receive any samples or photos for investigation.The device history records and retain samples could not be reviewed as the lot number is unknown.This complaint has not been confirmed for the indicated failure modes: defective locking mechanism and tube push off.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that when using one (1) bd vacutainer® eclipse¿ blood collection needle pre-attached holder, the safety mechanism separated, and tube push off from the non patient end of the cannula occurred.No adverse impact was reported regarding the patient or user.
|
| |
|