• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752180
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
A non-healthcare professional reported that water chamber had completed prime but loses suction during intraocular lens implant surgery.The surgery was successfully completed on same day.There was no information about the patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer's internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in h.6.And h.11.Product evaluation was not able to be performed since the reported product was not received at the investigation site for evaluation.Product is expected but has not been returned to date.Receipt of complaint product or additional information pertinent to this complaint will result in re-evaluation of the complaint investigation.The investigation conducted a non-conformance review of the reported lot number.No deviations that could have contributed to this event were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information, the investigation was unable to conclusively identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.No further investigative or manufacturing actions are warranted at this time due to the product not being returned to the manufacturing site for testing.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23838754
Report Number1644019-2025-05558
Device Sequence Number15834401
Product Code LRO
UDI-Device Identifier00380657521807
UDI-Public(01)00380657521807(17)270430(11)250508(10)17DY3R
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065752180
Device Lot Number17DY3R
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/24/2025
Supplement Date Received by Manufacturer01/26/2026
Initial Report FDA Received Date12/18/2025
Supplement Report FDA Received Date02/18/2026
Date Device Manufactured05/08/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-