Product complaint # (b)(4).Date sent to the fda: 12/18/2025 this report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to eth for evaluation.Visual inspection revelated that one empty opened foil packet and one winding former with a needle suture piece were received for analysis.Product code sxpp1a406.During visual inspection of the returned sample, marks that appear to be caused by a surgical instrument were observed on the body needle.The suture was examined, and the end of the suture was noted to be cut probably caused by a surgical instrument.In addition, crimping marks and body fluids were found on the strand.The other section of the suture was not returned for analysis.As the suture is absorbable and the duration of exposure of the suture in the environment could not be determined, the functional test could not be performed.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|