| Model Number |
720081-01 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient with this tactra malleable penile prosthesis was dissatisfied with the device, stating that it felt too short.A surgical procedure was performed in which the device was removed and replaced.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient with this tactra malleable penile prosthesis was dissatisfied with the device, stating that it felt too short.A surgical procedure was performed in which the device was removed and replaced.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to b1: adverse event/product problem.Upon receipt at our quality assurance laboratory, this tactra malleable penile prosthesis underwent a thorough analysis.Microscopic inspection revealed that the cylinders were cut to 23 cm with a sharp instrument.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of "length too short" was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available, the conclusion code of known inherent risk of device was assigned to this investigation.
|
| |
|
Search Alerts/Recalls
|