| Model Number |
DDU-100E |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.The cause of the aed not powering on was related to the customers failure to maintain the battery pack.As a follow up with the customer, the proper maintenance of this device was reviewed.
|
| |
|
Event or Problem Description
|
|
An international distributor reported that their customer's aed would not power on; however, when tested with a spare battery pack, it did power on and prompted a service code.The customer did not report this occurring during patient use.
|
| |
|