• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 CXR; PROSTHESIS, PENIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 CXR; PROSTHESIS, PENIS, INFLATABLE Back to Search Results
Model Number 72404014
Medical Device Problem Codes Inflation Problem (1310); Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Detailed product information was not provided to bsc.We completed a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available the cause that contributed to the reported event cannot be established.
 
Event or Problem Description
It was reported that the patient's inflatable penile prosthesis (ipp) exhibited inflation issues.During a surgical procedure, air was observed inside the device.The entire ipp was not removed; only the quick-connect coupling from an accessory kit was explanted.The patient's condition post-procedure was stable, and no complications were reported.
 
Event or Problem Description
It was reported that the patient's inflatable penile prosthesis (ipp) exhibited inflation issues.During a surgical procedure, air was observed inside the device.The entire ipp was not removed; only the quick-connect coupling from an accessory kit was explanted.The patient's condition post-procedure was stable, and no complications were reported.
 
Additional Manufacturer Narrative
Concomitant product udi for pump is (b)(4).Concomitant product udi for reservoir is (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available the cause that contributed to the reported event cannot be established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CXR
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key23840303
Report Number2124215-2025-91765
Device Sequence Number21290808
Product Code JCW
UDI-Device Identifier00878953002712
UDI-Public(01)00878953002712(17)290627(10)1100538316
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number72404014
Device Catalogue Number72404014
Device Lot Number1100538316
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2025
Supplement Date Received by Manufacturer12/18/2025
Initial Report FDA Received Date12/18/2025
Supplement Report FDA Received Date01/08/2026
Date Device Manufactured06/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PUMP: UPN: 72404310, LOT: 1100629059; RESERVOIR: UPN: 720182-01, LOT: 1100657025
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age40 YR
Patient SexMale
-
-