• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE, INC. CLOSURE TOP; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE, INC. CLOSURE TOP; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 3301-1
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2025-00381 through 3012447612-2025-00383.
 
Event or Problem Description
It was reported that three pathfinder closure tops were damaged when surgeon final torqued them: two screw¿s threads were broken, and one's inner hexagon stripped off.There was no patient impact; however, there was a 30 minute delay to prepare other closure tops.This is report two of three for this event.
 
Additional Manufacturer Narrative
Corrections in d9 and h3.Additional information in h6: investigation type, findings, and conclusions.A review of the dhr was conducted and found no indications of manufacturing issues.An inspection of the part shows the threads are fractured.This event likely cannot be attributed to manufacturing or design related issues.The complaint is confirmed for closure top for the failure of stripped thread.The failure is believed to be attributed to off-axis or excessive forces applied during use.This device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported that three pathfinder closure tops were damaged when surgeon final torqued them: two screw¿s threads were broken, and one's inner hexagon stripped off.There was no patient impact; however, there was a 30 minute delay to prepare other closure tops.This is report two of three for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLOSURE TOP
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
teresa george
10225 westmoor dr.
westminster, CO 80021
3035018506
MDR Report Key23840504
Report Number3012447612-2025-00382
Device Sequence Number9445521
Product Code NKB
UDI-Device Identifier00889024334861
UDI-Public(01)00889024334861(10)ABN
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K131980
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3301-1
Device Lot NumberABN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/18/2025
Supplement Report FDA Received Date03/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient EthnicityNon Hispanic
-
-