| Catalog Number |
0375638000 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the inner shaver blade cutting tip broke off when shaving.The metal piece was able to be retrieved.They opened up another shaver blade to get through case.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: tip broke off.Probable root cause: inadequate window profile.Inadequate welding process.Improper material strength.Inadequate size selected for the application.Material brittleness.Excessive force applied by the user.Incorrect rfid tag.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the inner shaver blade cutting tip broke off when shaving.The metal piece was able to be retrieved.They opened up another shaver blade to get through case.
|
| |
|
Search Alerts/Recalls
|