| Model Number |
DDU-120A |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Troubleshooting of the device with the customer identified that the aed was placed into service without the pads connected to the aed.The ddu series aeds are designed to be stored with the pads connector already installed.This simplifies the procedure for deploying and operating the device in an emergency.During the aed's automated self-test, the aed would have identified that the pads were not connected and attempted to alert the user by flashing its red asi and chirping.The aed would continue to chirp and flash until the aed's condition is addressed or the battery pack becomes completely depleted.The cause of the aed not powering on was related to a failure to set-up the aed and maintain the battery pack.
|
| |
|
Event or Problem Description
|
|
A customer reported that their aed's asi is flashing red, and the aed will not power on.The customer did not report this occurring during patient use.
|
| |
|