WESTMED LLC ATELLICA VTR SYRINGES; 3ML VENTED LUER LOCK SYRINGE, (~50IU)16.7IU/ML BALANCED HEPARIN, VENTED TIP CAP
|
Back to Search Results |
|
| Model Number |
11561345 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product involved in the report was discarded and is not available for return.A review of the device history record is in-progress.All information reasonably known as of 18 dec 2025 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
It was reported, blood leaked while expressing air, and splashed out from the top of the cap onto the operator's face.It was additionally reported, the operator was wearing gloves and goggles and followed the hospital guidelines for blood exposure.While investigating the event, it was found that the affected syringe was missing the white filter of the tip cap.There was no report of a delay in therapy, hurt, harm or injury as a result of this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: a1.Additional information-investigation completion: d4; h2; h5; h6.The actual complaint product was not available to be returned for evaluation; additionally, no photographic/videographic evidence was provided by the reporter.In the absence of a photo or video, the complaint could not confirmed; without a sample to perform functional evaluation, a root cause could not be determined.The device history record for lot 405075 was reviewed and the product was produced according to product specifications.All information reasonably known as of 26 feb 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by sun med holdings llc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to sun med holdings llc.Sun med holdings llc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the sun med holdings complaint database and identified as complaint-(b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an sun med holdings llc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
It was reported, blood leaked while expressing air, and splashed out from the top of the cap onto the operator's face.It was additionally reported, the operator was wearing gloves and goggles and followed the hospital guidelines for blood exposure.While investigating the event, it was found that the affected syringe was missing the white filter of the tip cap.There was no report of a delay in therapy, hurt, harm or injury as a result of this reported event.
|
| |
|