| Model Number |
DM0010FAA |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a burr hole procedure, three perforators broke while drilling.It was reported that procedure was delayed by 20 minutes and procedure was completed using a manual drill instead of the intended powered device. at the time of report, there was no known impact to a patient.Additional details regarding patient impact is being followed from the facility.Additional information was received stating that there was no patient or staff impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: no conclusion can be drawn.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis:evaluation couldn't able to reproduce the reported malfunction of perforator broke.The device worked as expected and no malfunctions or non conformities were identified.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|