MAUDE Adverse Event Report: MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOU Back to Search Results
Model Number DM0210FAA
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis: no conclusion can be drawn.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a burr hole procedure, three perforator broke while drilling.It was reported that procedure was delayed by 20 minutes and procedure was completed using a manual drill instead of the intended powered device.At the time of report, there was no known impact to a patient.Additional details regarding patient impact is being followed from the facility.Additional information was received stating that there was no patient or staff impact.
 
Additional Manufacturer Narrative
H3: product analysis:evaluation couldn't able to reproduce the reported malfunction of perforator broke.The device worked as expected and no malfunctions or non conformities were identified.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYDRILL PERFORATOR
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOU
Manufacturer (Section D)
MICROMAR INDUSTRIA E COMERCIO LTDA
cpnj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -390
BR  09980-390
Manufacturer (Section G)
MICROMAR INDUSTRIA E COMERCIO LTDA
cpnj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -390
BR   09980-390
Manufacturer Contact
natalia matos
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980--390
BR   09980-390
22713209
MDR Report Key23847014
Report Number1625507-2025-02400
Device Sequence Number21214099
Product Code HBF
UDI-Device Identifier07898959543609
UDI-Public(01)07898959543609(17)290510(10)1238/25
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K141455
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDM0210FAA
Device Catalogue NumberDM0210FAA
Device Lot Number1238/25
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2025
Supplement Date Received by Manufacturer03/27/2026
Initial Report FDA Received Date12/19/2025
Supplement Report FDA Received Date04/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/10/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
-
-