MAUDE Adverse Event Report: MICROVENTION, INC. V-TRAK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. V-TRAK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 7210-0515-A2
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that an embolization coil implant was deployed and detached in a coil mass of multiple coils.While they were placing the last coil, it detached unexpectedly with half of the coil length in the aneurysm and the other half left in the parent vessel.It was reported that the patient is fine and the surgeon is not concerned.The surgeon placed stents along side the portion of the coil that was still in the vessel to "pin it" so it would not migrate.
 
Additional Manufacturer Narrative
Based on a review of the device¿s risk documentation, the reported event did not indicate there were the presence of any potential or new manufacturing, design, quality, or other systemic issues, or non-conformance's.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform an analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformance's associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential complications potential complications include, but are not limited to: hematoma at the site of entry, vessel perforation, aneurysm rupture, parent artery occlusion, incomplete aneurysm filling, emboli, hemorrhage, ischemia, vasospasm, coil migration or misplacement, premature or difficult coil detachment, clot formation, revascularization, post-embolization syndrome, and neurological deficits including stroke and possibly death.Cases of chemical aseptic meningitis, edema, hydrocephalus and/or headaches have been associated with the use of embolization coils in the treatment of large and giant aneurysms.The physician should be aware of these complications and instruct patients when indicated.Appropriate patient management should be considered.Warnings and precautions the hes is intended for single use only.Do not resterilize and/or reuse the device.After use, dispose in accordance with hospital, administrative and/or local government policy.Do not use if the packaging is breached or damaged.¿ the hes must be delivered only through a wire-reinforced microcatheter with a ptfe inner surface coating.Damage to the device may occur and necessitate removal of both the hes and microcatheter from the patient.Do not advance the v trak delivery pusher with excessive force.Determine the cause of any unusual resistance, remove the hes and check for damage.Advance and retract the hes device slowly and smoothly.Remove the entire hes if excessive friction is noted.If excessive friction is noted with a second hes, check the microcatheter for damage or kinking.If repositioning is necessary, take special care to retract the coil under fluoroscopy in a one-to-one motion with the v trak delivery pusher.If the coil does not move in a one-to-one motion with the v trak delivery pusher, or if repositioning is difficult, the coil may have become stretched and could possibly break.Gently remove and discard the entire device.Due to the delicate nature of the hes coils, the tortuous vascular pathways that lead to certain aneurysms and vessels, and the varying morphologies of intracranial aneurysms, a coil may occasionally stretch while being maneuvered.Stretching is a precursor to potential coil breakage and migration.If resistance is encountered while withdrawing a coil that is at an acute angle relative to the microcatheter tip, it is possible to avoid coil stretching or breaking by carefully repositioning the distal tip of the catheter at, or slightly inside, the ostium of the aneurysm.By doing so, the aneurysm and artery act to funnel the coil back into the microcatheter.Introduction and deployment of the hes 19.Seat the distal tip of the introducer sheath at the distal end of the microcatheter hub and close the rhv lightly around the introducer sheath to secure the rhv to the introducer.Do not over-tighten the rhv around the introducer sheath.Excessive tightening could damage the device.20.Push the coil into the lumen of the microcatheter.Use caution to avoid catching the coil on the junction between the introducer sheath and the hub of the microcatheter.Initiate timing using a stopwatch or timer at the moment the device enters the microcatheter.Detachment must occur within the specified reposition time.21.Push the hes through the microcatheter until the proximal end of the v trak delivery pusher meets the proximal end of the introducer sheath.Loosen the rhv.Retract the introducer sheath just out of the rhv.Close the rhv around the v trak delivery pusher.Slide the introducer sheath completely off of the v trak delivery pusher.Use care not to kink the delivery system.To prevent premature hydration of the hes, ensure that there is flow from the saline flush.24.Under fluoroscopic guidance, slowly advance the hes coil out the tip of the microcatheter.Continue to advance the hes coil into the lesion until optimal deployment is achieved.Reposition if necessary.If the coil size is not suitable, remove and replace with another device.If undesirable movement of the coil is observed under fluoroscopy following placement and prior to detachment, remove the coil and replace with another more appropriately sized coil.Movement of the coil may indicate that the coil could migrate once it is detached.Do not rotate the v trak delivery pusher during or after delivery of the coil into the aneurysm.Rotating the hes v trak delivery pusher may result in a stretched coil or premature detachment of the coil from the v trak delivery pusher, which could result in coil migration.Angiographic assessment should also be performed prior to detachment to ensure that the coil mass is not protruding into the parent vessel.25.Complete the deployment and any repositioning so that the coil will be detached within the reposition time specified.After the specified time, the swelling of the hydrophilic polymer may prevent passage through the microcatheter and damage the coil.If the coil cannot be properly positioned and detached within the specified time, simultaneously remove the device and the microcatheter.The device was implanted in the patient and was not available for evaluation to investigate if a condition existed that would have contributed to the reported event.Supplemental imaging was also not provided for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation further examine the cause of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-TRAK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key23847124
Report Number2032493-2025-90726
Device Sequence Number10030369
Product Code KRD
UDI-Device Identifier00816777023230
UDI-Public00816777023230
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K091641
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7210-0515-A2
Device Catalogue Number7210-0515
Device Lot Number0000791293
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2025
Initial Report FDA Received Date12/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age88 YR
Patient SexFemale
-
-