• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. SAFSITE®; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. SAFSITE®; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 415068
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the user facility: event 4: reason of complaint: customer states: "we had 4 incidences of the safsite bungs leaking - 3 occurred with nuclear medicine technologists and one with a registered nurse.Patients were cannulated, the bung attached to the end of the cannula and the cannula taped down.Once the cap was removed from the safsite bung so a flush could be administered, blood started to leak back through the one-way valve of the bung.Bungs were secured tightly we are always vigilant with this as we are injecting radioactive pharmaceuticals through these, so it was not a case of a loose connection.It wasn't just a trickle back through the one-way valve; it was a significant flow.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation, the root cause of the reported incident was unable to be determined.The reported defect was unable to be confirmed.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Additionally, the retained units were evaluated and passed the internal testing.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFSITE®
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
824 12th ave
bethlehem PA 18018
Manufacturer (Section G)
B. BRAUN MEDICAL INC.
824 12th ave
bethlehem PA 18018
Manufacturer Contact
elena reifenstein
824 12th ave
bethlehem, PA 18018
8334251464
MDR Report Key23848936
Report Number2521402-2025-01584
Device Sequence Number9958198
Product Code FPA
UDI-Device Identifier04022495769862
UDI-Public(01)04022495769862(11)241105(17)291031(10)0061968348
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K942988
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number415068
Device Lot Number0061968348
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2025
Supplement Date Received by Manufacturer12/09/2025
Initial Report FDA Received Date12/19/2025
Supplement Report FDA Received Date01/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-