ARROW INTERNATIONAL LLC ARROW REDIGUARD IAB: 7FR 30CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAB-S730C
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# (b)(4).
 
Event or Problem Description
It was reported "after the catheter inserted into the patient, the pump frequent alarms high pressure.After the catheter removed, the front one-third of the balloon failed to inflate.Change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
 
Event or Problem Description
It was reported "after the catheter inserted into the patient, the pump frequent alarms high pressure.After the catheter removed, the front one-third of the balloon failed to inflate.Change the new catheter".No patient injury or consequence reported.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The serial number on the returned sample is (b)(6).Returned for investigation was a 30cc 7.0fr rediguard intra-aortic balloon catheter (iabc) with the original packaging box that matches the serial number of the returned sample.The sample was returned in a cardboard box and was loosely packed within the original packaging tray/carton.Upon return, the one-way valve was connected and tethered to the short driveline tubing.The bladder was fully unwrapped.Spots of dried blood were noted on the exterior surfaces of the returned sample.No obvious blood was noted within the helium pathway.No visual damage or abnormalities were noted to the returned sample.The bladder thickness was measured at six points with measurements ranging from 0.0058in-0.0068in and was within specification of process document.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute according to quality system document.An attempt to aspirate and flush the catheter using a 60cc lab-inventory syringe was u nable to be successfully completed due to a blocked/clotted central lumen.Immediate push back on the syringe plunger was experienced.The blockage, most likely dried blood, was unable to be cleared.The iabc was connected to an iabp with a lab in-ventory 30cc inflation driveline tubing.The iabc was inserted into the "t" tube and 75mmhg backpressure was applied.The iabc was pumped for a minimum of 15 minutes.There were no alarms triggered.The bladder inflated and deflated completely with each beat.The iabc was leak tested in accordance with testing methods from manufacturing procedure.No leaks were detect-ed.Full inflation was achieved.The device passed the leak test.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.The guidewire could not advance at approximately 36cm from the iabc distal tip, which is the location of the clotted central lumen.Dried blood was noted on the guidewire upon removal.The guidewire was front loaded through the iabc luer.The guidewire could not advance at approximately 29.5cm from the iabc luer, which is the location of the clotted central lumen.Dried blood was noted on the guidewire upon removal.A device histo-ry record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing sp ecifications prior to release.The reported complaint that the "front one-third of the balloon failed to inflate" is not confirmed.The returned iabc bladder was fully intact with no abnormalities noted.During the investigation, the iabc fully inflated, and no leaks/alarms were noted.The returned device passed visual and functional test specifications.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further ac-tion required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW REDIGUARD IAB: 7FR 30CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
sachin sony
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key23850088
Report Number3010532612-2025-01251
Device Sequence Number9958713
Product Code DSP
UDI-Device Identifier10801902145649
UDI-Public(01)10801902145649(11)241108(17)261031(10)18F24L0005(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K981660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberIAB-S730C
Device Lot Number18F24L0005
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2025
Supplement Date Received by Manufacturer01/08/2026
Initial Report FDA Received Date12/19/2025
Supplement Report FDA Received Date01/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-