| Catalog Number |
07251254190 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The tacrolimus reagent expiration date was not provided.The cobas e 801 module serial number was (b)(6) the qc was acceptable.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
We received an allegation about discrepant results for 6 patients' samples tested with elecsys tacrolimus assay on a cobas e 801 module.Sample 1: initial result: >40 ug/l.Repeat result: 11.3 ug/l.Sample 2: initial result: >40 ug/l.Repeat result: 10.7 ug/l.Sample 3: initial result: >40 ug/l.Repeat result: 10.1 ug/l.Sample 4: initial result: >40 ug/l.Repeat result: 11.6 ug/l.Sample 5: initial result: >40 ug/l.Repeat result: 12.8 ug/l.Sample 6: initial result: >40 ug/l.Repeat result: 15.7 ug/l.The physician questioned the initial results and requested that the samples be repeated.The repeat results were deemed to be correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The measuring cell was exchanged on (b)(6) 2025.A general reagent issue can be excluded since the qc was within the ranges on the day of the event.The field service engineer performed a liquid flow cleaning.No further issues were reported after the service visit.The investigation did not identify a product problem.The cause of the event was consistent with a non-optimal customer instrument maintenance.
|
| |
|
Search Alerts/Recalls
|