| Catalog Number |
B5LT |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, the obturator shaft was returned bent.The obturator tip was fractured and the tip was not returned for investigation.The device inspection indicated that the complaint was confirmed for the reported failure mode.A review of the dhr supports that the device met all inspections and test criteria prior to release from stryker.The reported event could be attributed to: - excessive force applied - contact with hard object or other improper handling - insufficient structural integrity - shipping/handling damage or extreme conditions the instructions for use (ifu) state: - damage to the instrument can lead to patient injuries.Always inspect instrument carefully for overall integrity before use.- do not use excessive force.- careful handling of instruments is necessary to avoid damage or breakage.- inspect the instruments for any damage.Do not use the instrument if any damage is noted.Return the instrument and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.- a comprehensive understanding of the principles and techniques involved in laser, electrosurgical, and ultrasonic procedures is essential to avoid shock and burn hazards to both patient and medical personnel and damage to the device or other medical instruments.The reported event will continue to be monitored through post-market surveillance.
|
| |
|
Event or Problem Description
|
|
It was reported that the 5mm plastic trocar tip broke off.An x-ray was taken with no evidence of presence in the abdomen.The patient went from pacu to ct.As reported, the patient had to have a ct scan post-op.There was no other patient injury or medical intervention reported and there was no extended procedure time reported.These are commonly used devices that are readily available.
|
| |
|
Search Alerts/Recalls
|