|
The initial reporter stated they received discrepant glucose results for one patient tested with accuchek inform ii meter serial numbers (b)(6).At 15:18, a sample from the patient was tested with meter serial number (b)(6), resulting in a glucose value of 46 mg/dl.At 15:20, a sample from the patient was tested with meter serial number (b)(6), resulting in a glucose value of 90 mg/dl.At 15:24, a sample from the patient was tested with meter serial number (b)(6), resulting in a glucose value of 56 mg/dl.At 15:24, a sample from the patient was tested with meter serial number (b)(6), resulting in a glucose value of 101 mg/dl.The customer believed the value of 101 mg/dl to be correct.
|