| Model Number |
11500A |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Fatigue (1849)
|
| Date of Event |
11/25/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 21mm 11500a aortic valve underwent a valve-in-valve procedure after an implant duration of 7 years, 3 months due to moderate stenosis.The patient presented with dyspnea and fatigue.The tavr was performed with a 23mm 9755rsl transcatheter valve.The patient was stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b4, g3, g6, h2.H11: corrected data: b3.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b4, b7, g3, g6, h2, h6 (type of investigation, investigation findings, investigation conclusions).Stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including hyperlipidemia (hld), and coronary artery disease (cad).The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.
|
| |
|
Search Alerts/Recalls
|