Model Number 1040005
Medical Device Problem Code Failure to Calibrate (2440)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to the manufacturer for service.There was no serious patient harm or injury that occurred or was reported.The device was not in patient use.During the evaluation of the device at the manufacturer's service center, the device failed the positive flow verification test step during testing.The device was recalibrated to address the issue.The device was retested and failed the dp purge valves test step during testing.The sensor board needs replaced to address the issue.Additionally, unrelated to the test failures, the device was found to have missing/displace rubber feet and the porting block was damaged.Rubber feet and the porting block needs to be replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 200 VENTILATOR, U.S.A.
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23851727
Report Number2518422-2025-112902
Device Sequence Number13154751
Product Code CBK
UDI-Device Identifier00606959022737
UDI-Public(01)00606959022737
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K093416
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1040005
Device Catalogue Number1040005
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/16/2025
Initial Report FDA Received Date12/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/15/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-