• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON UNITY CATARACT FMS PROCEDURE PACK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON UNITY CATARACT FMS PROCEDURE PACK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065000339
Medical Device Problem Code Suction Failure (4039)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A customer reported that the products failed to aspirate, so the surgeon opted for another machine.Additional information has been requested but none received till date.
 
Additional Manufacturer Narrative
Additional information provided in h.6.And h.11.The product was not received at the investigating site for this complaint report; visual inspection or functional testing could not be conducted.The investigation conducted a non-conformance review of the reported lot number.No relevant deviations were identified; all corresponding production release specifications defined in the device master record were met.As the root cause and its origin are inconclusive, further investigation or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.No further investigative or manufacturing actions are warranted at this time due to the product not being returned to the manufacturing site for testing.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNITY CATARACT FMS PROCEDURE PACK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23852121
Report Number1644019-2025-05593
Device Sequence Number18534843
Product Code LRO
UDI-Device Identifier00380650003393
UDI-Public(01)00380650003393(17)270228(11)250326(10)17ATLW
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065000339
Device Lot Number17ATLW
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/26/2025
Supplement Date Received by Manufacturer02/13/2026
Initial Report FDA Received Date12/19/2025
Supplement Report FDA Received Date03/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/26/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-