Catalog Number 623131015
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/03/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional manufacturer narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during implantation the stemless component on the humeral frame and handle on the stemless component fractured the anterior humerus leading to use of a stemmed component.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary- no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot- the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: b1 (adverse event), b2 (required intervention) corrected: d4 (primary udi number), h1 (type of reportable event), h6 (health effect - clinical code, health effect - impact code, medical device problem code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTACT HUMERAL HANDLE
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23853922
Report Number1818910-2025-22297
Device Sequence Number20081029
Product Code LXH
UDI-Device Identifier10603295559627
UDI-Public(01)10603295559627(10)EZ253053
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number623131015
Device Lot NumberEZ253053
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/08/2025
Supplement Date Received by Manufacturer12/29/2025
Initial Report FDA Received Date12/19/2025
Supplement Report FDA Received Date12/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/20/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-