ALCON RESEARCH, LLC - HUNTINGTON CLAREON TORIC UV ABSORBING IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number CCW0T3
Medical Device Problem Code Inaccurate Delivery (2339)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that during an intraocular lens (iol) implant procedure, there was a lens delivery issue.The lens was fully inserted and removed during the same procedure.The surgery was completed on the same day with another lens.Additional information has been requested.
 
Additional Manufacturer Narrative
Corrected information provided in d.10.Correction: on initial mdr the concomitant medical product, monarch iol delivery system, cartridge was an error.Additional information has been provided in d.10., h.3., h.6., h.11.The product was not returned.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.The account indicated the use of a cartridge and handpiece that are not qualified for use with this lens.A viscoelastic was indicated which is qualified for use with this lens with the use of qualified lens/cartridge/handpiece combinations.The root cause is most likely related to a failure to follow the ifu.The account used an unqualified lens/cartridge/handpiece combination.Company foldable iols are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON TORIC UV ABSORBING IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23854047
Report Number1119421-2025-03187
Device Sequence Number13297919
Product Code MJP
UDI-Device Identifier00380652431903
UDI-Public(01)00380652431903(17)270827(11)230829(10)15703302
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCCW0T3
Device Lot Number15703302
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/26/2025
Supplement Date Received by Manufacturer01/13/2026
Initial Report FDA Received Date12/19/2025
Supplement Report FDA Received Date02/06/2026
Date Device Manufactured08/29/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC VISCOELASTIC SYSTEM.; MONARCH IOL DELIVERY SYSTEM, CARTRIDGE.; NON-COMPANY CARTRIDGE.; NON-COMPANY PLATINUM 1 INJECTOR.; UNKNOWN INJECTOR/HANDPIECE.; UNKNOWN VISCOELASTIC.
Patient Age81 YR
Patient SexUnknown
-
-