| Model Number |
620-0221511-01 |
| Medical Device Problem Codes |
Pacing Problem (1439); Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device displayed a "pace defib device failure" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was confirmed to be caused by a defective processor/bridge/pace (pbp) board.The pbp board was replaced to remedy the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|