DEPUY IRELAND - 3015516266 EMSYS LNR IMP TIP 40; HIP INSTRUMENT - IMPACTOR Back to Search Results
Catalog Number 481440050
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional manufacturer narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unspecified surgical procedure it was observed that the impactor tip broke when impacting the liner.No additional information could be provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary.Impactor tip broke when impacting the liner.The product was not returned to depuy synthes, however photos were provided for review.See attachment img_7714.Png, img_7715.Png, img_7716.Png.The photo investigation revealed that the emsys lnr imp tip 40 is broken in multiple pieces.All fracture features observed are consistent with overload fracture.The fracture initiated and progressed due to overload, whereby a load was applied exceeding the ultimate strength of the material.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the emsys lnr imp tip 40 would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summary: impactor tip broke when impacting the liner.The product was returned to depuy synthes for evaluation.Visual inspection of the returned device found that the emsys lnr imp tip was broken around the threaded section.Broken fragments were not returned for evaluation.All fracture features observed are consistent with overload fracture.The fracture initiated and progressed due to overload, whereby a load was applied exceeding the ultimate strength of the material.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the emsys lnr imp tip 40 would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Corrected: h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) investigation summary: impactor tip broke when impacting the liner.The product was not returned to depuy synthes; however photos were provided for review.See attachment (b)(4).The photo investigation revealed that the emsys lnr imp tip 40 is broken in multiple pieces.All fracture features observed are consistent with overload fracture.The fracture initiated and progressed due to overload, whereby a load was applied exceeding the ultimate strength of the material.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the emsys lnr imp tip 40 would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: d9.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMSYS LNR IMP TIP 40
Common Device Name
HIP INSTRUMENT - IMPACTOR
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23854333
Report Number1818910-2025-22265
Device Sequence Number10374922
Product Code HWA
UDI-Device Identifier10603295510208
UDI-Public(01)10603295510208(10)AZ639178
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number481440050
Device Lot NumberAZ639178
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/10/2025
Supplement Date Received by Manufacturer01/14/2026
02/05/2026
02/24/2026
Initial Report FDA Received Date12/19/2025
Supplement Report FDA Received Date01/15/2026
02/06/2026
02/26/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/05/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-